inforMED
MalfunctionHTQ

ACTIS BROACH SZ 1

Received Mar 11, 2026 · Event occurred Feb 26, 2026

Report 1818910-2026-04601 · MDR key 24575576

Device

Generic name

Broaches

Manufacturer

Depuy Ireland

Catalog number

201001010

Lot number

PG318200

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: D4 (LOT), D9, H4. CORRECTED: D4 (PRIMARY UDI NUMBER). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: IT WAS REPORTED THAT THIS FACILITY HAD MULTIPLE VERSION 1 AND 2 BROACHES BREAK AND WERE REQUESTING ALL OF THEIR CONSIGNED BROACHES BE SWAPPED OUT FOR VERSION 3 TO AVOID FURTHER BREAKAGES. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL ANALYSIS OF THE ACTIS BROACH REVEALED THAT THE DEVICE EXHIBITS AN OVERALL WORN APPEARANCE CONSISTENT WITH NORMAL AND CONSTANT USAGE. NO DEFECTS WERE OBSERVED A DIMENSIONAL INSPECTION AND FUNCTIONAL TESTING WERE NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE ACTIS BROACH WOULD NOT CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. CORRECTED: H3.

Description of Event or Problem

IT WAS REPORTED THE BROACH WAS BROKEN.

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.