inforMED
MalfunctionPRI

ELECSYS BRAHMS PCT

Received Mar 11, 2026 · Event occurred Feb 16, 2026

Report 1823260-2026-00863 · MDR key 24570434

Device

Generic name

Procalcitonin Assay

Manufacturer

Roche Diagnostics

Catalog number

08828679190

Lot number

87572901

Product problems

  • Incorrect Measurement
  • High Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF A QUESTIONABLE ELECSYS BRAHMS PCT RESULT FROM THE COBAS E 801 ANALYTICAL UNIT FOR ONE PATIENT. THE INITIAL RESULT WAS 0.69 NG/ML. THE REPEAT RESULT WAS 0.05 NG/ML, ACCOMPANIED BY A DATA FLAG. THE REPEAT RESULT WAS DEEMED TO BE CORRECT. THE SAMPLE WAS REPEATED AFTER THERE WAS A QC FAILURE FOR ELECSYS BRAHMS PCT.