inforMED
MalfunctionFMZ

GIRAFFE INCUBATOR CARESTATION

Received Mar 11, 2026 · Event occurred Feb 12, 2026

Report 3036470679-2026-00027 · MDR key 24569122

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SOLENOID SUPPLY VALVE ASSEMBLY WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.