HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°
Received Mar 9, 2026 · Event occurred Feb 17, 2026
Report 3007960282-2026-00042 · MDR key 24553492
Device
Generic name
Hf-resection Electrode
Manufacturer
Olympus Medical Products Czech, Spol. S R.o.Model number
WA22302DLot number
1000222534
Product problems
- Break
- Detachment of Device or Device Component
Patient
NA · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING A THERAPEUTIC HYSTEROSCOPY, THE RESECTION ELECTRODE BROKE DURING USE IN THE PATIENT (INTRAUTERINE). THE FRACTURED FRAGMENT WAS RETRIEVED USING BARREL FORCEPS. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE WITH A DELAY OF LESS THAN 30 MINUTES. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.
Additional Manufacturer Narrative
THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
Additional Manufacturer Narrative
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D9, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS CONFIRMED. THE ELECTRODE LOOP HAD BROKEN OFF COMPLETELY. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTION WAS IDENTIFIED DURING THE DEVICE EVALUATION: INSULATION HOSES WERE DAMAGED DUE TO THERMAL OVERLOAD. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE COMPLAINT IS NOT ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.