TOTALCARE BARIATRIC PLUS BED

Received Mar 9, 2026 · Event occurred Feb 11, 2026

Report 1824206-2026-00530 · MDR key 24547434

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P1840D

Catalog number

P1840DCA000043

Product problems

  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BAXTER TECHNICIAN FOUND THE 4 CASTERS NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL, THE TOTALCARE BARIATRIC BED AND TOTALCARE BARIATRIC PLUS THERAPY SYSTEM REQUIRE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM SEMIANNUAL PREVENTIVE MAINTENANCE (PM). PM WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. INSPECT THE CASTERS FOR PROPER MOUNTING, EXCESSIVE WEAR, OR TREAD DAMAGE. REPLACE AS NEEDED. VERIFY PROPER OPERATION OF THE BRAKE SYSTEM. ADJUST AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN FEB 14, 2025. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE 4 CASTERS TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem

ON 11-FEB-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT TOTALCARE BARIATRIC CAPITAL (PRODUCT CODE P1840DCA000043, SERIAL NUMBER (B)(6)), HAD CASTERS MOVE WHEN BRAKE IS ENGAGED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.