VEGA® RV IMPLANT Ø4.0 X 14MM
Received Mar 9, 2026 · Event occurred Mar 2, 2026
Report 3003267134-2026-00002 · MDR key 24545620
Device
Product problems
- Fracture
Patient
67 YR · Unknown
- Difficulty Chewing
Narrative
Additional Manufacturer Narrative
MECHANICAL FAILURE DUE TO IMPLANT FATIGUE CAUSED BY OVERLOAD, EXACERBATED BY CRESTAL BONE LOSS, IS ALSO EVIDENT. IT IS DIFFICULT TO SPECULATE ON THE CAUSE, AS NO INFORMATION IS AVAILABLE ON THE MORPHOLOGY OF THE RESTORED CROWN, ALTHOUGH IT CAN BE CONFIRMED THAT THE IMPLANT IS LOCATED IN AN AREA SUBJECT TO HIGH MECHANICAL STRESS. FRACTURE OF THIS TYPE OF DEVICE IS A POSSIBLE BUT UNLIKELY MECHANICAL COMPLICATION, WITH A RELATIVELY LOW INCIDENCE WHEN THE INSTRUCTIONS FOR USE ARE FOLLOWED (EQUAL OR LESS THAN 1% ACCORDING TO THE VALUES REPORTED IN THE CLINICAL EVALUATION OF KLOCKNER® DEVICES). HOWEVER, THE COMBINATION OF ALL THE FACTORS MENTIONED IN THE TECHNICAL CAUSES INCREASES THE PROBABILITY OF ITS OCCURRENCE.
Description of Event or Problem
THE CUSTOMER REPORTS THE BREAKAGE OF A VEGA RV IMPLANT IN A PATIENT. THEY PROVIDE THE CORRESPONDING QUALITY REPORT AND X-RAYS OF THE IMPLANT PLACED WITHOUT THE PROSTHESES.
Remedial action
- Notification