Received Mar 6, 2026 · Event occurred Feb 4, 2026
Report 3003950207-2026-00005 · MDR key 24536370
Device
Generic name
Reliance 6500 Endoscope Drying And Storage Cabinet
Manufacturer
Steris CorporationModel number
RELIANCE 6500Catalog number
REL650020EM
Product problems
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT THE LEFT DOOR OF THEIR RELIANCE 6500 ENDOSCOPE DRYING AND STORAGE CABINET FELL OFF. NO REPORT OF INJURY.
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE 6500 SERIES RELIANCE ENDOSCOPE DRYING AND STORAGE CABINET AND FOUND THAT THE BOTTOM PIN THAT SECURES THE DOOR WAS LOOSE RESULTING IN THE REPORTED EVENT. THE TECHNICIAN REINSTALLED THE DOOR WITH A REPLACEMENT PIN. THE UNIT WAS TESTED, CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.