inforMED
MalfunctionFMZ

GIRAFFE OMNIBED CARESTATION

Received Mar 6, 2026 · Event occurred Feb 8, 2026

Report 3008729547-2026-00020 · MDR key 24533113

Device

Generic name

Incubator, Neonatal

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • Insufficient Flow or Under Infusion

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SERVO OXYGEN PCB WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.