inforMED
MalfunctionFLL

DEROYAL

Received Mar 5, 2026 · Event occurred Feb 7, 2026

Report 24518823 · MDR key 24518823

Device

Generic name

Thermometer, Electronic, Clinical

Model number

81-020409

Catalog number

020409

Lot number

62530402

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ESOPHAGEAL TEMP PROBE PLACED FOR CORE TEMP MONITORING WITH WHOLE BODY COOLING PROTOCOL. PLUGGED ESOPHAGEAL TEMP PROBE INTO CORD CONNECTED TO BLANKETEROL AND THE MACHINE DID NOT RECOGNIZE A TEMPERATURE AND ALARMED CHECK PROBE. BLANKETEROL TEMP CORD PLUGGED INTO RECTAL TEMP PROBE AND WORKED. AFTER CONFIRMING THAT THE CORD FROM THE BLANKETEROL WAS WORKING TRIED TO USE WITH ESOPHAGEAL TEMP PROBE AGAIN BUT IT WOULD NOT READ A TEMPERATURE. REMOVED ESOPHAGEAL PROBE AND PLACED A NEW ONE. NEW ESOPHAGEAL PROBE READING WITHOUT ISSUE.