inforMED
MalfunctionPZI

STATSTRIP GLUCOSE HOSPITAL METER SYSTEM

Received Mar 3, 2026 · Event occurred Feb 1, 2026

Report 1219029-2026-00014 · MDR key 24500277

Device

Generic name

Prescription Use Blood Glucose Meter For Near-patient Testing

Model number

2.0

Catalog number

63685

Product problems

  • Computer Software Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE IN QUESTION IS A STATSTRIP GEN2 GLUCOSE METER. HOWEVER, THE SPECIFIC SERIAL NUMBER WAS NOT PROVIDED. TECHNICAL SUPPORT WAS ABLE TO MANUALLY TRANSMIT THE STUCK RESULT, WHICH REMOVED THE BACKLOG OF RESULTS. AN INVESTIGATION INTO THE ROOT CAUSE OF THE REPORTED ISSUE HAS BEGUN BUT IS NOT YET COMPLETED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem

RESULT FROM STATSTRIP GEN 2 METER TRANSMITTED TO CUSTOMER'S MIDDLEWARE STUCK IN "PENDING" STATUS WHICH PREVENTED OTHER RESULTS FROM CROSSING TO PATIENTS CHART. ALTHOUGH THERE WERE NO SPECIFIC ALLEGATIONS OF DELAYS IN TREATMENT, THIS ISSUE COULD POTENTIALLY LEAD TO DELAYS.