AQUABEAM ROBOTIC SYSTEM
Received Mar 2, 2026 · Event occurred Jan 21, 2026
Report 3012977056-2026-00035 · MDR key 24491294
Device
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Male
- Hemorrhage/Blood Loss/Bleeding
Narrative
Additional Manufacturer Narrative
THE AQUABEAM ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE IT IS CURRENTLY IN THE POSSESSION OF THE USER FACILITY. THE INVESTIGATION OF THIS EVENT CONSISTED OF A REVIEW OF THE DEVICE HISTORY RECORD (DHR) AND INSTRUCTIONS FOR USE (IFU). A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR AB2000/SERIAL NUMBER (B)(6) WAS CONDUCTED, WHICH CONFIRMED THAT THERE WERE NO NON-CONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE SYSTEM MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION. A REVIEW OF THE LOG FILES FOR THIS PROCEDURE WAS NOT CONDUCTED AS IT WOULD NOT PROVIDED ANY ADDITIONAL INFORMATION. THE AQUABEAM ROBOTIC SYSTEM'S INSTRUCTION FOR USE (IFU), SJ-IFU0101-00, REV. C, WAS REVIEWED. 4.3 WARNINGS: PROCEDURE ¿ AS WITH ANY SURGICAL UROLOGIC PROCEDURE, POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING, SOME OF WHICH MAY LEAD TO SERIOUS OUTCOMES AND MAY REQUIRE INTERVENTION: BLEEDING. 8.33 STERILE: A. AFTER THE AQUABEAM HANDPIECE REMOVAL, FOLLOW THE STANDARD CLOT EVACUATION PROCEDURE TO REMOVE CLOTS AND TISSUE WITH A CYSTOSCOPIC SHEATH BY USING AN ELLIK BLADDER EVACUATOR OR TOOMEY SYRINGE. B. USE ONE OF THE FOLLOWING METHODS TO ACHIEVE HEMOSTASIS: ¿ CAUTERY FOLLOWED BY FOLEY BALLOON CATHETER INSERTION. ¿ UNDER SPINAL ANESTHESIA, INSERT A BALLOON CATHETER IN THE BLADDER WITH BLADDER NECK TRACTION THEN FILL THE BLADDER WITH STERILE SALINE AND MAINTAIN FOR APPROXIMATELY 30-60 MINUTES BEFORE STARTING CBI (CONTINUOUS BLADDER IRRIGATION) ¿ BALLOON CATHETER IN BLADDER WITH BLADDER NECK TRACTION ¿ BALLOON CATHETER IN PROSTATIC FOSSA: ¿ INFLATE BALLOON WITH 5CC IN THE BLADDER ¿ UNDER TRUS GUIDANCE RETRACT BALLOON INTO PROSTATIC FOSSA ¿ INFLATE BALLOON TO 30-50% OF INITIAL PROSTATE VOLUME ¿ APPLY MILD TRACTION ON THE CATHETER TO HOLD THE BALLOON CATHETER IN PLACE ¿ BALLOON CATHETER IN BLADDER, NO TRACTION NOTE: CONSIDERATIONS FOR CAUTERY: > START AT BLADDER NECK. > PERFORM FOCAL CAUTERY AT SOURCES OF BLEEDING. SOURCES OF BLEEDING MAY BE COVERED UP BY RESIDUAL TISSUE ABLATED BY THE WATERJET ("FLUFFY TISSUE"). IN ORDER TO CHECK FOR ALL SOURCES OF BLEEDING, FIRST US THE LOOP TO REMOVE FLUFFY TISSUE. > TURN OFF IRRIGATION AND INSPECT FOR SOURCES OF BLEEDING UNDER NORMAL BLOOD PRESSURE. C. START CBI PER HOSPITAL PROTOCOL. DO NOT ALLOW IRRIGATION FLUID BAGS TO BECOME EMPTY. MONITOR EFFLUENT COLOR. THE AQUABEAM ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. THE AQUABEAM ROBOTIC SYSTEM'S IFU LISTS BLEEDING AS A POTENTIAL RISK OF AQUABLATION THERAPY AND PROVIDES ADEQUATE INSTRUCTIONS TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT THE PATIENT FELL WHICH CAUSED THE CATHETER TO DARKEN AND HEMOGLOBIN DROPPED. THE PATIENT WAS THEN TRANSFUSED AND STAYED IN THE HOSPITAL FOR AN EXTENDED TIME. THE PATIENT HAS SINCE BEEN DISCHARGED. BASED ON THE EVENT DETAILS, PLUS A REVIEW OF THE DHR AND IFU, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS (PROCEPT) BECAME AWARE THAT, POST-AQUABLATION THERAPY, THE PATIENT FELL IN THE POST-ANESTHESIA CARE UNIT (PACU). IT WAS NOTED THAT THE PATIENT'S CATHETER DARKENED, AND HEMOGLOBIN LEVELS DROPPED. THE PATIENT WAS THEN TRANSFUSED AND STAYED IN THE HOSPITAL FOR AN EXTENDED TIME. THE PATIENT HAS SINCE BEEN DISCHARGED. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.