DEPTH GAUGE FOR LCKNG SCRS TO 100
Received Feb 26, 2026 · Event occurred Feb 12, 2026
Report 8030965-2026-01967 · MDR key 24457084
Device
Product problems
- Material Deformation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D10. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: CORRECTED: G1. H3, H6: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE TIP OF THE DEPTH GAUGE FOR LCKNG SCRS TO 100 WAS DEFORMED. THE OBSERVED CONDITION WAS CONSISTENT AS AN END OF LIFE INDICATOR FOR THE DEVICE. NO OTHER ISSUES WERE IDENTIFIED. THE OBSERVED CONDITION WAS IDENTIFIED AS AN END OF LIFE INDICATOR; DAMAGE CONSISTENT WITH REPEATED USE AND SERVICING. THE LIFECYCLE REQUIREMENTS OF THE DEVICE ARE EVENT RELATED AND DEPEND ON THE USE AND INSPECTION OF THE DEVICE IN CLINICAL PRACTICE. AS THE DEVICE CAN BE DAMAGED ON THE FIRST OR 100TH USE, THE DEVICE MUST BE PROPERLY INSPECTED PRIOR TO EACH SURGICAL USE. REFER TO THE DEVICE/COUNTRY SPECIFIC IFU FOR INFORMATION RELATED TO END OF LIFE, REPROCESSING INSTRUCTIONS, AND INSPECTION PROCEDURES. A DIMENSIONAL INSPECTION WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. A FUNCTIONAL EVALUATION WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF DEPTH GAUGE FOR LCKNG SCRS TO 100 WOULD HAVE CONTRIBUTED TO THE COMPLAINED ISSUE. IT WAS DETERMINED THAT THE REUSABLE INSTRUMENT WAS WORN FROM REPEATED USE AND SERVICING. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE IS TRACED TO END OF LIFE, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS REQUIRED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: PART: 03.010.428, LOT: L952004, MANUFACTURING SITE: WERK BALSTHAL, SUPPLIER: NA, RELEASE TO WAREHOUSE DATE: 07 AUG 2018, EXPIRATION DATE; NA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D4 (LOT, EXPIRATION DATE), D9, H4. CORRECTED: D4 PRIMARY UDI NUMBER. THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON (B)(6) 2026 THERE WAS AN ISSUE IN PRE-OP WHEN SETTING UP FOR A TIBIAL NAIL. THE AIMING SHOWED SIGNS OF WEAR ON THE LEVER HANDLE, IT HAD A CRACK IN IT. THE DEPTH GAUGE THAT WAS USED WAS ALSO BENT AND COULD NOT BE STRAIGHTENED. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
PRODUCT COMPLAINT (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.