FISHER & PAYKEL HEALTHCARE

Received Feb 24, 2026

Report 9611451-2026-00712 · MDR key 24438661

Device

Generic name

Infant Warmer

Model number

IW990AGU

Catalog number

IW990AGU

Lot number

031126

Product problems

  • Crack

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

DURING DEVICE EVALUATION AND PERFORMANCE TESTING OF A IW990AGU WALL MOUNT INFANT WARMER AT OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTRE IN GERMANY, IT WAS FOUND THAT THE WARMER HEAD WAS DAMAGED. THERE WAS NO REPORTED PATIENT INVOLVEMENT

Additional Manufacturer Narrative

(B)(4). SECTION D4: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. SECTION G4: THE IW990AGU MANUAL-CONTROL WALL MOUNT INFANT WARMER IS NOT SOLD IN THE UNITED STATES OF AMERICA (USA) BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K971695. METHOD: THE COMPLAINT IW990AGU WALL MOUNT INFANT WARMER WAS RECEIVED AT OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTRE IN GERMANY WHERE IT WAS VISUALLY INSPECTED AND SERVICED BY A TRAINED F&P SERVICE TECHNICIAN. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE F&P SERVICE TECHNICIAN, AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: VISUAL INSPECTION OF THE SUBJECT DEVICE REVEALED THAT THE WARMER HEAD WAS DAMAGED. CONCLUSION: WE ARE UNABLE TO DETERMINE THE EXACT CAUSE OF THE OBSERVED DAMAGE TO THE IW990AGU WALL MOUNT INFANT WARMER. THE USER MANUAL FOR THE IW990 INFANT WARMER STATES THE FOLLOWING: CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYDROCHLORITES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE. CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES. THE DEVICE TECHNICAL/SERVICE MANUAL ALSO CONTAINS A CHECKLIST WHICH SPECIFIES THAT USERS PERFORM SAFETY, PERFORMANCE, AND FUNCTIONAL CHECKS AT LEAST ONCE A YEAR.