inforMED
MalfunctionJQC

EXPRESS 4 CENTRIFUGE

Received Feb 24, 2026 · Event occurred Jun 5, 2024

Report 2050012-2026-00013 · MDR key 24437443

Device

Generic name

Centrifuges (micro, Ultra, Refrigerated) For Clinical Use

Catalog number

X00-005531-001

Lot number

N/A

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THE STATSPIN EXPRESS 4 CENTRIFUGE¿S ROTOR SHATTERED, PRODUCING SMOKE AND A BURNING SMELL. THE CUSTOMER CONFIRMED NO EXPOSURE, INJURIES OR DEATH WERE ASSOCIATED WITH THIS EVENT.

Additional Manufacturer Narrative

TROUBLESHOOTING, CONDUCTED BY THE CUSTOMER AND A BECKMAN CUSTOMER TECHNICAL SPECIALIST (CTS), CONFIRMED THE EXPRESS 4 ROTOR WAS BROKEN. THE CUSTOMER REPORTED THE LID REMAINED CLOSED AND NO PARTS ESCAPED THE CENTRIFUGE DURING THE INCIDENT. CTS RECOMMENDED CUSTOMER REPAIR OR REPLACEMENT OF THE ROTOR TO RESOLVE THE ISSUE. SECTION A2, A3, A4, AND A5: INFORMATION NOT PROVIDED BY CUSTOMER. BECKMAN COULTER INTERNAL IDENTIFIER IS (B)(4).

Additional Manufacturer Narrative

THE FOLLOWING SECTIONS WERE UPDATED: B1: CHANGED FROM PRODUCT MALFUNCTION TO A NOT REPORTABLE EVENT. H10: FOLLOWING ADDITIONAL REVIEW AND CONSULTATION WITH THE BECKMAN MEDICAL DIRECTOR ON FEBRUARY 25, 2026. IT WAS DETERMINED THAT THE NON-PATIENT SAFETY RELATED RISK FOR EXPOSURE TO PHYSICAL AND MECHANICAL ENERGY HAZARDS HAS AS A SEVERITY OF MINOR DAMAGE AND A PROBABILITY OF HARM (P2) AS OCCASIONAL. BASED ON THIS RISK ASSESSMENT, THE EVENT IS NOT REPORTABLE.