GIRAFFE INCUBATOR
Received Feb 21, 2026 · Event occurred Jan 26, 2026
Report 3036470679-2026-00011 · MDR key 24408885
Device
Product problems
- No Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SERVO MODULE WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. D4. PRIMARY UDI NUMBER: THERE IS NO UDI AVAILABLE AS THE DEVICE WAS MANUFACTURED PRIOR TO UDI COMPLIANCE REQUIREMENTS. BLOCK H4: DATE OF DEVICE MANUFACTURE YEAR IS 2003. THE MONTH OF MANUFACTURE WAS UNAVAILABLE AT TIME OF MDR FILING. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.