MalfunctionFMZ

GIRAFFE INCUBATOR

Received Feb 21, 2026 · Event occurred Jan 26, 2026

Report 3036470679-2026-00011 · MDR key 24408885

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Catalog number

NA

Lot number

NA

Product problems

  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SERVO MODULE WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. D4. PRIMARY UDI NUMBER: THERE IS NO UDI AVAILABLE AS THE DEVICE WAS MANUFACTURED PRIOR TO UDI COMPLIANCE REQUIREMENTS. BLOCK H4: DATE OF DEVICE MANUFACTURE YEAR IS 2003. THE MONTH OF MANUFACTURE WAS UNAVAILABLE AT TIME OF MDR FILING. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.