inforMED
MalfunctionBTQ

TRACHEAL TUBE

Received Feb 20, 2026 · Event occurred Jan 14, 2026

Report 8040412-2026-00032 · MDR key 24399908

Device

Generic name

Tube, Tracheal (w/wo Connector

Model number

IPN040737

Catalog number

100380-000025

Lot number

40E24C2436

Product problems

  • Failure to Disconnect
  • Physical Resistance/Sticking

Patient

NA · Unknown

  • Insufficient Information

Narrative

Description of Event or Problem

IT WAS REPORTED THAT: "CONNECTOR WAS FIXED TO THE TUBE, COULDN'T BE DETACHED." ASSOCIATED COMPLAINTS 8040412-2026-00031 AND 8040412- 2026-00032.

Additional Manufacturer Narrative

(B)(4).

Additional Manufacturer Narrative

(B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS NO SAMPLE WAS RETURNED FOR ANALYSIS. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED, AND NO RELEVANT FINDINGS WERE IDENTIFIED. WITHOUT THE DEVICE TO EVALUATE, THE COMPLAINT COULD NOT BE CONFIRMED, AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.

Description of Event or Problem

IT WAS REPORTED THAT: "CONNECTOR WAS FIXED TO THE TUBE, COULDN'T BE DETACHED." ASSOCIATED COMPLAINTS 8040412-2026-00031 AND 8040412- 2026-00032.