inforMED
MalfunctionODC

GAS/WATER VALVE

Received Feb 20, 2026

Report 9610595-2026-16265 · MDR key 24399002

Device

Generic name

Gas/water Valve

Model number

MAJ-2010

Product problems

  • Microbial Contamination of Device

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE EVALUATION OF THE EVENT IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT THE GAS/WATER VALVE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

Additional Manufacturer Narrative

UPDATED FIELDS: D4,D6,D8,E1,H3 THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE COMPLETED INVESTIGATION'S REVIEW OF THE CUSTOMER CLEANING DISINFECTION AND STERILIZATION (CDS) PROCESSES, RESULTS OF THIRD-PARTY TESTING, THE DEVICE EVALUATION AND THE FINAL INVESTIGATION. THE COMPLAINT INVESTIGATION REVIEWED THE CUSTOMER PROVIDED CDS PROCESSES WHERE NO OBVIOUS DEVIATIONS FROM INSTRUCTIONS FOR USE (IFU) WERE IDENTIFIED. THE USER FACILITY PROVIDED THE FOLLOWING RESULT OF THE CULTURE TEST, PERFORMED AT THE THIRD-PARTY LABS: SAMPLING DATE: JAN 29, 2026 SAMPLING FROM: ENTIRE BUTTON CFU: 6.8 BACTERIAL IDENTIFICATION: CLECSIELLA GENUS BASED ON THE RESULTS OF THE INVESTIGATION AND BECAUSE THE SUBJECT DEVICE WAS NOT RETURNED, THE REPORTED EVENT COULD NOT BE CONFIRMED AND A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem

NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.