inforMED
MalfunctionJJE

AU5800 CHEMISTRY ANALYZER

Received Feb 19, 2026 · Event occurred Feb 12, 2026

Report 9612296-2026-00079 · MDR key 24394326

Device

Generic name

Analyzer, Chemistry (photometric, Discrete), For Clinical Use

Catalog number

B96698

Product problems

  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE AU5800 CHEMISTRY ANALYZER AND IDENTIFIED THE PROBABLE CAUSE AS A DEFECTIVE REFERENCE VALVE AND THE REFERENCE ELECTRODE. THE FSE REPLACED THE REFERENCE VALVE AND THE REFERENCE ELECTRODE. TO RESOLVE THE ISSUE. SECTION A2, A3, A4, A5 AND A6: INFORMATION NOT PROVIDED BY CUSTOMER. THE BECKMAN COULTER INTERNAL IDENTIFIER IS (B)(4).

Description of Event or Problem

THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW SODIUM PATIENT RESULTS ON THEIR AU5800 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.