RADIOLUCENT HOHMANN RETRACTOR W/18 L240
Received Feb 19, 2026 · Event occurred Jan 1, 2025
Report 8030965-2026-01665 · MDR key 24388933
Device
Product problems
- Material Integrity Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THE ISSUE HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT#: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THE TIP SURFACE WAS DEFORMED BY DEEP SCRATCHES AND SMALL BROKEN APART FRAGMENTS. BROKEN FRAGMENTS WERE NOT RETURNED. NO OTHER ISSUE WAS IDENTIFIED. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A COMPONENT FAILURE THAT WAS CAUSED BY EXPOSURE TO UNINTENDED FORCES WITH OTHER TOOLS AND HARD EDGES COMING. PROPERLY HANDLING AND ATTENTION TO THE APPROVED USE OF THE DEVICE DIMINISHES THE RISK OF FAILURE. A FUNCTIONAL TEST WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. A DIMENSIONAL INSPECTION WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE RADIOLUCENT HOHMANN RETRACTOR W/18 L240 WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DIMENSIONAL INSPECTION: N/A. DEVICE HISTORY: PART#: 03.100.109, SYNTHES LOT#: J011037, SUPPLIER LOT#: J011037, RELEASE TO WAREHOUSE DATE: 17 AUG 2022, SUPPLIER: (B)(4), NO NCRS WERE GENERATED DURING PRODUCTION.