MalfunctionCGN

ELECSYS FOLATE III

Received Feb 19, 2026 · Event occurred Feb 2, 2026

Report 1823260-2026-00591 · MDR key 24388215

Device

Generic name

Acid, Folic, Radioimmunoassay

Manufacturer

Roche Diagnostics

Catalog number

08324174190

Lot number

91120301

Product problems

  • Failure to Recalibrate
  • High Test Results
  • High Readings
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE E 402 ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS FOLATE III ON A COBAS PURE E 402 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN A FOLATE VALUE OF > 20.0 NG/ML. THE SAMPLE WAS AUTOMATICALLY DILUTED 1:2 AND REPEATED, RESULTING IN A VALUE OF 12.4 NG/ML. THE SAMPLE WAS REPEATED AGAIN WITHOUT DILUTION, RESULTING IN A VALUE OF > 20.0 NG/ML. THE SAMPLE WAS REPEATED WITH AN ON-BOARD 1:2 DILUTION, RESULTING IN A FOLATE VALUE OF 11.0 NG/ML ON (B)(6) 2026. THE SAMPLE WAS MANUALLY DILUTED 1:2 AND REPEATED, RESULTING IN A RAW VALUE OF 6.29 NG/ML, WHICH CALCULATES TO A FINAL FOLATE VALUE OF 12.58 NG/ML ON (B)(6) 2026. A NEW REAGENT PACK WAS CALIBRATED AND THE SAMPLE WAS REPEATED WITHOUT DILUTION ON (B)(6) 2026, RESULTING IN A FOLATE VALUE OF 9.52 NG/ML. THE SAMPLE WAS ALSO DILUTED 1:2 ON-BOARD THE INSTRUMENT AND REPEATED, RESULTING IN A VALUE OF 8.05 NG/ML ON (B)(6) 2026. THE SAMPLE WAS REPEATED AGAIN USING THE OLDER REAGENT PACK AND IT RESULTED IN A FOLATE VALUE OF 17 NG/ML.