inforMED
MalfunctionLNH

INGENIA

Received Feb 19, 2026 · Event occurred Feb 16, 2026

Report 3003768277-2026-100804 · MDR key 24386469

Device

Generic name

Full-body Mri System, Superconducting Magnet

Catalog number

781260

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS DAMAGED. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

Additional Manufacturer Narrative

THE QBC SEAL THAT WAS DAMAGED WAS REPLACED AND THE SYSTEM WAS HANDED BACK TO THE CUSTOMER FOR USE. THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION. REMEDIAL ACTION: A FIELD ACTION WAS INITIATED UNDER 2023-PD-MR-013. A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE. THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DECEMBER 2024.

Remedial action

  • Modification/Adjustment