ALLY ADAPTIVE CATARACT TREATMENT SYSTEM

Received Feb 18, 2026 · Event occurred Feb 3, 2026

Report 3009026057-2026-00005 · MDR key 24377055

Device

Generic name

Ally Adaptive Cataract Treatment System

Manufacturer

Lensar, Inc.

Model number

N/A

Catalog number

70-00050-001

Lot number

N/A

Product problems

  • Patient-Device Incompatibility

Patient

NA · Unknown

  • Corneal Perforation

Narrative

Description of Event or Problem

ON (B)(6) 2026, DOCTOR ((B)(4)) REPORTED TO CAS THAT THEY HAD A FULL THICKNESS AK ON PROCEDURE ID #(B)(6).

Additional Manufacturer Narrative

REVIEW OF PROCEDURE #(B)(6) SCHEIMPFLUG IMAGING NOTES BUBBLE ON EDGE OF SECOND AK. LASER ADJUSTED TO ACCOMMODATE THE THINKER CORNEA. SURGEON MAY CONSIDER CHANGING SETTINGS TO 80-85% DEPTH INSTEAD OF A FIXED 600 MICRON DEPTH TO MINIMIZE RISK OF THIN POSTERIOR CORNEA POST ARCUATE. SYSTEM FUNCTIONED AS DESIGNED. NO FURTHER CLINICAL FOLLOW-UP REQUIRED.