inforMED
MalfunctionFMZ

GIRAFFE INCUBATOR

Received Feb 18, 2026 · Event occurred Jan 21, 2026

Report 3036470679-2026-00006 · MDR key 24376811

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Catalog number

MA

Lot number

MA

Product problems

  • Insufficient Flow or Under Infusion

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SERVO WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: PATIENT INFORMATION COULD NOT BE OBTAINED DUE TO COUNTRY PRIVACY LAWS. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN SIGNIFICANT DEGRADATION OF OXYGEN THERAPY DURING A CASE. THERE WAS NO PATIENT INJURY.