MEDEX 3-WAY STOPCOCK EXTENSION SET

Received Feb 17, 2026 · Event occurred May 5, 2025

Report 1526863-2026-00021 · MDR key 24368482

Device

Generic name

Tubing, Fluid Delivery

Catalog number

MX43660

Lot number

NI

Product problems

  • Break
  • Fluid/Blood Leak

Patient

12 MO · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H6: CODES WERE UPDATED. ONE DEVICE WAS RETURNED FOR INVESTIGATION. THE REPORTED COMPLAINT SEPARATION WAS CONFIRMED. DURING VISUAL INSPECTION, THE PRESSURE TUBING WAS RECEIVED SEPARATED FROM THE FEMALE LUER OF THE STOPCOCK. WHEN THE END OF THE TUBING WAS MICROSCOPICALLY EXAMINED, THERE WAS GOOD SOLVENT COVERAGE AROUND THE TUBING. THE PROBABLE CAUSE OF THE SEPARATION IS UNKNOWN.

Additional Manufacturer Narrative

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE LINE REPEATEDLY BROKE, RESULTING IN BLOOD LOSS. THE EVENT OCCURRED DURING INFUSION, MONITORING, AND INTERVENTIONS, AND REQUIRED ADDITIONAL MONITORING AND INTERVENTIONS. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM WAS REPORTED.