MI Z HANDLE ACET REAMER
Received Feb 13, 2026 · Event occurred Jan 23, 2026
Report 1020279-2026-00215 · MDR key 24333838
Device
Generic name
Prosthesis, Hip, Femoral Component, Cemented, Metal
Manufacturer
Smith & Nephew, Inc.Catalog number
71364073
Lot number
24GM09509
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT, DURING A THR SURGERY ONE (1) MI Z HANDLE ACET REAMER AND ONE (1) R3 OFFSET IMPACTOR TIP BROKE ON SURGEON WHILE HE WAS REAMING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Additional Manufacturer Narrative
INTERNAL COMPLAINT REFERENCE (B)(4). H11 H3, H6: THIS COMPLAINT WAS OPENED BY SMITH+NEPHEW TO DOCUMENT A PRODUCT PROBLEM ASSOCIATED WITH A SMITH+NEPHEW DEVICE. THE REPORTED PROBLEM RELATES TO KNOWN RISKS WITH THE DEVICE ITSELF OR WITH ITS USE THAT ARE APPROPRIATELY DOCUMENTED IN OUR RISK FILES. SMITH+NEPHEW WILL CONTINUE TO MONITOR TRENDS IN ACCORDANCE WITH OUR POST-MARKET SURVEILLANCE PROCESS AND TAKE NECESSARY ACTION AS REQUIRED IF ANTICIPATED SEVERITY AND/OR OCCURRENCE RATES ARE EXCEEDED. SMITH+NEPHEW HAS NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY SPECIFICATIONS AT THE TIME OF MANUFACTURE. BASED ON OUR REVIEW OF ALL CURRENTLY AVAILABLE INFORMATION, WE ARE UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. HOWEVER, AS THE USE OF OUR PRODUCT CANNOT BE EXCLUDED AS A POTENTIAL CAUSE OR CONTRIBUTORY FACTOR TO THE REPORTED ISSUE, WE ARE SUBMITTING THIS REPORT IN ACCORDANCE WITH APPLICABLE REGULATIONS. IF ADDITIONAL INVESTIGATIVE FINDINGS OR INFORMATION BECOMES AVAILABLE THAT ALTERS THE CONCLUSIONS OF THIS REPORT, A FOLLOW-UP REPORT WILL BE SUBMITTED AS REQUIRED.
Additional Manufacturer Narrative
H3, H6: THE ASSOCIATED DEVICE WAS RETURNED AND EVALUATED. THE VISUAL INSPECTION REVEALED THE DEVICE RETURNED WITH THE PROXIMAL JOINT FRACTURED. THE ENTIRE DEVICE IS WORN FROM USE WITH THE IDENTIFICATION MARKINGS NEARLY ILLEGIBLE. A REVIEW OF THE PRODUCTION ORDER DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. A REVIEW OF COMPLAINT HISTORY REVEALED SIMILAR EVENTS FOR THE LISTED PART NUMBER OVER THE PREVIOUS 12 MONTHS, BUT NO SIMILAR EVENTS FOR THE BATCH BASED ON THE HISTORICAL DATA, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. A HISTORICAL REVIEW CONCLUDED THAT A SIMILAR EVENT WAS PREVIOUSLY IDENTIFIED, AND ESCALATED ACTIONS WERE PERFORMED. IT HAS BEEN CONCLUDED THAT THERE IS A POTENTIAL FOR BREAKAGE IF USED AFTER THE EXPECTED LIFETIME OR EXCESSIVE FORCE IS USED AS THIS DEVICE IS A REUSABLE INSTRUMENT AND IT IS EXPECTED TO WEAR OVER TIME. ALSO, THE FAILURE RATE OF THIS ISSUE IS WITHIN EXPECTED AND DOCUMENTED IN THE RISK FILE. AT THIS TIME, WE HAVE NO EVIDENCE TO CONCLUDE THAT THE PRODUCT FAILED TO MEET ANY SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY POTENTIAL FACTORS THAT COULD CONTRIBUTE TO THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.