Received Feb 12, 2026 · Event occurred Jan 13, 2026

Report 3005899764-2026-00011 · MDR key 24326067

Device

Generic name

Sterilizer

Catalog number

SR0201110111

Product problems

  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT THEIR AMSCO 400 STEAM STERILIZER WAS LEAKING WATER ONTO THE FLOOR. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE STERILIZER AND FOUND THAT THE STERILIZER'S GENERATOR PIPING REQUIRED ADJUSTMENT. WHILE ONSITE THE TECHNICIAN ADJUSTED THE GENERATOR PIPING, PROACTIVELY REPLACED THE FLOW REGULATOR AND CHECK VALVE, TESTED THE UNIT, AND CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION. THE STERILIZER WAS RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.