ATELLICA CH URINE ALBUMIN (UALB)
Received Feb 12, 2026 · Event occurred Jan 3, 2026
Report 2432235-2026-00046 · MDR key 24323281
Device
Generic name
Albumin, Antigen, Antiserum, Control
Manufacturer
Siemens Healthcare Diagnostics Inc.Catalog number
11537225
Lot number
150472
Product problems
- Low Test Results
Patient
77 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN OUTSIDE OF UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING ERRONEOUSLY DEPRESSED URINE ALBUMIN (UALB) RESULT OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. SIEMENS HEALTHINEERS HAS CONFIRMED A POTENTIAL FOR FALSELY DEPRESSED RESULTS FOR A SMALL SUBSET OF SAMPLES ABOVE THE MEASURING INTERVAL WHEN USING THE ATELLICA CH URINE ALBUMIN (UALB) ASSAY ON THE ATELLICA CH AND ATELLICA CI ANALYZERS. US CUSTOMERS WERE NOTIFIED THROUGH AN URGENT MEDICAL DEVICE CORRECTION (UMDC, LETTER ACHC26-03.A.US) AND OUS CUSTOMERS WERE IDENTIFIED THROUGH URGENT FIELD SAFETY NOTICE (UFSN, LETTER ACHC26-03.A.OUS) TITLED ¿ATELLICA CH URINE ALBUMIN (UALB) ¿ POTENTIAL FOR FALSELY DEPRESSED RESULTS¿. THE COMMUNICATION WAS DIRECTED TO ALL CUSTOMERS WHO RECEIVED ATELLICA CH URINE ALBUMIN (UALB) IMPACTED LOTS INFORMING THEM OF THE ISSUE AND PROVIDING INSTRUCTIONS THE CUSTOMER MUST FOLLOW.
Description of Event or Problem
THE CUSTOMER REPORTED TO SIEMENS AN ERRONEOUSLY DEPRESSED URINE ALBUMIN (UALB) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPROCESSED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED URINE ALBUMIN (UALB) RESULT.
Remedial action
- Recall