inforMED
MalfunctionDCF

ATELLICA CH URINE ALBUMIN (UALB)

Received Feb 12, 2026 · Event occurred Dec 16, 2025

Report 2432235-2026-00044 · MDR key 24320435

Device

Generic name

Albumin, Antigen, Antiserum, Control

Model number

N/A

Catalog number

11537225

Lot number

150472

Product problems

  • Low Test Results

Patient

37 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN OUTSIDE THE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS CUSTOMER CARE CENTER (CCC) REGARDING AN ERRONEOUSLY DEPRESSED URINE ALBUMIN (UALB) RESULT OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. SIEMENS HEALTHINEERS HAS CONFIRMED A POTENTIAL FOR FALSELY DEPRESSED RESULTS FOR A SMALL SUBSET OF SAMPLES ABOVE THE MEASURING INTERVAL WHEN USING THE ATELLICA CH URINE ALBUMIN (UALB) ASSAY ON THE ATELLICA CH AND ATELLICA CI ANALYZERS. US CUSTOMERS WERE NOTIFIED THROUGH AN URGENT MEDICAL DEVICE CORRECTION (UMDC, LETTER ACHC26-03.A.US) AND OUS CUSTOMERS WERE IDENTIFIED THROUGH URGENT FIELD SAFETY NOTICE (UFSN, LETTER ACHC26-03.A.OUS) TITLED ¿ATELLICA CH URINE ALBUMIN (UALB), POTENTIAL FOR FALSELY DEPRESSED RESULTS¿. THE COMMUNICATION WAS DIRECTED TO ALL CUSTOMERS WHO RECEIVED ATELLICA CH URINE ALBUMIN (UALB) IMPACTED LOTS INFORMING THEM OF THE ISSUE AND PROVIDING INSTRUCTIONS THE CUSTOMER MUST FOLLOW.

Description of Event or Problem

THE CUSTOMER REPORTED TO SIEMENS AN ERRONEOUSLY DEPRESSED URINE ALBUMIN (UALB) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED DILUTED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE HIGHER RESULT OBTAINED WAS CONSIDERED CORRECT. THE SAME SAMPLE WAS REPROCESSED DILUTED ON THE ORIGINAL ATELLICA CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED UALB RESULT.

Remedial action

  • Recall