MalfunctionCEC

ALINITY I FREE T4 REAGENT KIT

Received Feb 11, 2026 · Event occurred Jan 27, 2026

Report 3005094123-2026-00066 · MDR key 24316193

Device

Generic name

Radioimmunoassay, Free Thyroxine

Catalog number

07P70-30

Lot number

82597UD00

Product problems

  • High Test Results

Patient

75 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS PROVIDED, NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE.

Description of Event or Problem

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE 75-YEAR-OLD FEMALE PATIENT. THE SAMPLE WAS REPEATED WITH A LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): ALINITY SERIAL NUMBER (B)(6) RESULT = 27.48 PMOL/L REPEAT RESULT = 17.94 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.84 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.16 PMOL/L. REFERENCE RANGE IS 9-19 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING DATA IDENTIFIED AN INCREASE IN COMPLAINTS FOR THE ALINITY I FREE T4 ASSAY AS WELL AS AN INCREASE IN COMPLAINT ACTIVITY FOR REAGENT LOT 82597UD00. HOWEVER, TESTING WAS PERFORMED UTILIZING RETAINED KITS OF LOT 82597UD00 AND NOTED THAT ALL TESTING PASSED, INDICATING THE REAGENT LOTS ARE PERFORMING AS EXPECTED. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT ASSAY AND COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE. PER THE LIMITATIONS OF THE PROCEDURE SECTION OF THE PACKAGE INSERT, RESULTS SHOULD BE USED IN CONJUNCTION WITH OTHER DATA, E.G., SYMPTOMS, RESULTS OF OTHER THYROID TESTS, CLINICAL IMPRESSIONS, ETC. IF THE FREE T4 RESULTS ARE INCONSISTENT WITH CLINICAL EVIDENCE, ADDITIONAL TESTING IS SUGGESTED TO CONFIRM THE RESULT. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 ASSAY, LOT 82597UD00 WAS IDENTIFIED.

Description of Event or Problem

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE 75-YEAR-OLD FEMALE PATIENT. THE SAMPLE WAS REPEATED WITH A LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID(B)(6): ALINITY SERIAL NUMBER (B)(6) RESULT = 27.48 PMOL/L REPEAT RESULT = 17.94 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.84 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.16 PMOL/L. REFERENCE RANGE IS 9-19 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.