inforMED
MalfunctionFOS

ARGYLE

Received Feb 10, 2026 · Event occurred Feb 4, 2026

Report 1423537-2026-00033 · MDR key 24310720

Device

Generic name

Catheter, Umbilical Artery

Model number

8888160556

Catalog number

8888160556

Lot number

2435200127

Product problems

  • Connection Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED. SECTION E1: E-MAIL: (B)(6).

Description of Event or Problem

THE CUSTOMER REPORTED THAT ARGYLE POLYURETHANE UMBILICAL VESSEL CATHETERS ARE NOT STAYING CONNECTED TO THE STOPCOCKS. IT OCCURS WITH EVERY ONE OF THEM THAT THEY HAVE OPENED. THE CONNECTOR ENDS DO NOT LOCK, THE TABS SEEM TO BE SLIPPING IN THE LUERLOCK OF THE STOPCOCKS. THEY JUST KEEP SPINNING, AND IF THEY APPLY ANY PULLING PRESSURE, THEY CAN PULL THEM RIGHT OUT. THERE WAS NO HARM REPORTED.

Additional Manufacturer Narrative

THE DEVICE HISTORY RECORD (DHR) FOR LOT 2435200127 WAS REVIEWED AND NO ANOMALIES RELATED TO THE REPORTED ISSUE WERE OBSERVED. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE REPORTED CONDITION WAS OBSERVED. A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO FURTHER INVESTIGATE THIS ISSUE.