InjuryHTW
LEVEL ONE
Received Feb 10, 2026 · Event occurred Jan 21, 2026
Report 9610905-2026-00007 · MDR key 24306763
Device
Product problems
- Fracture
Patient
25 YR · Male
- Device Embedded In Tissue or Plaque
Narrative
Description of Event or Problem
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Additional Manufacturer Narrative
THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU.