PRECICE® INTRAMEDULLARY LIMB LENGTHENING SYSTEM
Received Feb 9, 2026 · Event occurred Jan 12, 2026
Report 3006179046-2026-00102 · MDR key 24295164
Device
Generic name
Femur Intramedullary Nail
Manufacturer
Nuvasive Specialized Orthopedics, Inc.Model number
P10.7-80B275Lot number
4061916
Product problems
- Positioning Failure
Patient
45 YR · Male
- Failure of Implant
Narrative
Additional Manufacturer Narrative
THE DEVICE HAS NOT YET BEEN RECEIVED FOR EVALUATION. THE ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL INFORMATION.
Description of Event or Problem
IT WAS REPORTED THAT DURING A SURGERY TO RECHARGE THE PRECICE NAIL NEITHER THE FAST DISTRACTOR MAX NOR THE ERC WERE ABLE TO RETRACT THE NAIL DESPITE NUMEROUS ATTEMPTS. THE ERC SHOWED THERE WAS A CONNECTION BETWEEN MAGNETS WITH A GREEN CHECK MARK BUT THERE WAS NO ACTUAL MOVEMENT OF THE NAIL. THE NAIL WAS THEN REMOVED AND TESTED ON THE BACK TABLE WITH NO SUCCESS. THE NAIL HAD BEEN EXTRACTED TO 78 MM PRIOR TO THE RECHARGE SURGERY. POST-OP, THE PATIENT REPORTED THAT DURING HIS FINAL DISTRACTIONS WITH THE ERC VIA THE PRESCRIPTION HE WAS HEARING A CLUNKING SOUND. HIS PRESCRIBED TREATMENT WAS .33 MM X 3 TIMES A DAY.