inforMED
MalfunctionPSZ

BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF FLU A+B LABORATORY KIT

Received Feb 6, 2026 · Event occurred Jan 12, 2026

Report 3006948883-2026-00074 · MDR key 24283392

Device

Generic name

Devices Detecting Influenza A, B, And C Virus Antigens

Catalog number

256041

Lot number

UNKNOWN

Product problems

  • False Negative Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA/510(K)#: K121797;K132256;K132693;K133138;K151301;K152874;K160164. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED WHILE USING THE BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF FLU A+B LABORATORY KIT, AN UNSPECIFIED NUMBER OF FLU A FALSE NEGATIVE RESULTS WERE OBTAINED. IT WAS NOTED THE CUSTOMER WAS EXPECTING FLU A POSITIVE RESULTS. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED. THE CUSTOMER HAS NOT RESPONDED TO MULTIPLE FOLLOW UP ATTEMPTS BY BD FOR ADDITIONAL INFORMATION.

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: THIS STATEMENT SUMMARIZES THE INVESTIGATION RESULTS REGARDING A COMPLAINT THAT ALLEGES FALSE NEGATIVE WHEN USING KIT FLU A+B 30 TEST HOSPITAL VERITOR (MATERIAL#: 256041), BATCH NUMBER UNKNOWN. THE CUSTOMER REPORTED THAT THEY RECEIVED FALSE FLU A NEGATIVE RESULTS. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE ALL COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, VISUAL INSPECTION OF RETENTION SAMPLES, AND VISUAL INSPECTION OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. BATCH HISTORY RECORD (BHR) REVIEW AND RETAIN SAMPLE TESTING WERE NOT PERFORMED AS THERE WAS NO BATCH NUMBER PROVIDED. NO PHOTOS OR SAMPLES WERE RECEIVED; THEREFORE, RETURN SAMPLE ANALYSIS COULD NOT BE PERFORMED. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. A TREND ANALYSIS FOR FALSE NEGATIVE WAS CONDUCTED; NO ADVERSE TREND WAS IDENTIFIED. THERE WERE NO CORRECTIVE ACTIONS TAKEN AT THIS TIME.

Description of Event or Problem

IT WAS REPORTED WHILE USING THE BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF FLU A+B LABORATORY KIT, AN UNSPECIFIED NUMBER OF FLU A FALSE NEGATIVE RESULTS WERE OBTAINED. IT WAS NOTED THE CUSTOMER WAS EXPECTING FLU A POSITIVE RESULTS. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED. THE CUSTOMER HAS NOT RESPONDED TO MULTIPLE FOLLOW UP ATTEMPTS BY BD FOR ADDITIONAL INFORMATION.