ATELLICA CH ENZYMATIC HEMOGLOBIN A1C (A1C_E)
Received Feb 6, 2026 · Event occurred Jan 26, 2026
Report 2432235-2026-00030 · MDR key 24279378
Device
Generic name
Assay, Glycosylated Hemoglobin
Manufacturer
Siemens Healthcare Diagnostics, Inc.Model number
N/ACatalog number
11097536
Lot number
150453
Product problems
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN OUTSIDE OF THE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REGIONAL SUPPORT CENTER (RSC) REGARDING ERRONEOUSLY ELEVATED ENZYMATIC HEMOGLOBIN A1C (A1C_E) RESULTS OBTAINED FOR THREE (3) PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING THE EVENT.
Description of Event or Problem
THE CUSTOMER REPORTED TO SIEMENS ERRONEOUSLY ELEVATED ENZYMATIC HEMOGLOBIN A1C (A1C_E) RESULTS WERE OBTAINED ON THREE (3) PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN AND QUESTIONED. THE SAME PATIENT SAMPLES WERE REPROCESSED ON EITHER THE SAME OR AN ALTERNATE ATELLICA CH 930 ANALYZER. LOWER RESULTS WERE OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED A1C_E RESULTS.