inforMED
MalfunctionJIL

CLINITEK ADVANTUS

Received Feb 5, 2026 · Event occurred Jan 6, 2026

Report 3002637618-2026-00005 · MDR key 24268152

Add to watch list
Product code (JIL)
Manufacturer (Siemens Healthcare Diagnostics Inc.)
Model (10484765)

Device

Generic name

Clinitek Advantus

Catalog number

10484765

Product problems

  • False Negative Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SIEMENS REQUESTED MATERIALS TO BE RETURNED FOR FURTHER INVESTIGATION. THE CUSTOMER STATED THAT THE REAGENT WAS NO LONGER AVAILABLE AND THAT THE INSTRUMENT HAD BEEN SENT FOR SERVICING. SIEMENS HAS REQUESTED THE SERVICE REPORT. THE CAUSE OF THIS EVENT IS UNKNOWN.

Description of Event or Problem

THE CUSTOMER REPORTED THAT THEY RECEIVED A FALSE NEGATIVE LEUKOCYTE RESULT ON ONE PATIENT COMPARED TO RETESTING OF THE SAME SAMPLE ON THE SAME INSTRUMENT AND MICROSCOPIC ANALYSIS. THE EXACT MICROSCOPIC RESULTS WERE NOT PROVIDED BY THE CUSTOMER. THE CUSTOMER STATED THAT THEY PERFORMED MAINTENANCE AS REQUIRED AND RUNS QC DAILY AND IT WAS PASSING. THERE IS NO REPORT OF HARM DUE TO THIS EVENT.

Additional Manufacturer Narrative

SIEMENS HAS COMPLETED THE INVESTIGATION. THE RETURNED DEVICE WAS EVALUATED. TESTING WAS PERFORMED ON RETURNED DEVICE. THE DEVICE PASSED ALL FUNCTIONAL TESTS. REVIEW OF INTERNAL RECORDS DO NOT IDENTIFY ANY PRODUCTION ISSUES, DEVIATIONS, OR OTHER ESCALATIONS THAT WOULD BE RELATED TO THE CUSTOMER¿S REPORTED ISSUE. NO OTHER CUSTOMER COMPLAINTS WERE RECEIVED ON THE LOT IN QUESTION. THE CAUSE OF THE EVENT IS UNKNOWN. NO PRODUCT PROBLEM IDENTIFIED.