KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLAN
Received Feb 3, 2026 · Event occurred Jan 14, 2026
Report 1220246-2026-00388 · MDR key 24251816
Device
Generic name
Washer, Bolt, Nut
Manufacturer
Arthrex, Inc.Model number
KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLANCatalog number
AR-2372BLO
Lot number
15478356
Product problems
- Positioning Failure
- Physical Resistance/Sticking
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-2372BLO KNOTLESS AC TIGHTROPE OPEN REPAIR IMPLANT WITH PEC BUTTON WOULD NOT TENSION AND FAILED TO ACHIEVE AC JOINT REDUCTION. THE SURGEON ULTIMATELY CUT AND REMOVED THE IMPLANT AND TIED THE GRAFT OVER THE CONSTRUCT. THE CASE WAS DELAYED APPROXIMATELY FIVE MINUTES, WITH NO ADDITIONAL ANESTHESIA ADMINISTERED. THE ISSUE OCCURRED DURING A CLAVICLE ORIF AND AC JOINT STABILIZATION PROCEDURE WITH INTERNAL OPEN REDUCTION PERFORMED ON (B)(6) 2026. NO PATIENT HARM WAS REPORTED.
Additional Manufacturer Narrative
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: G3, H3, H6. BASED ON THE INFORMATION PROVIDED WHICH MAY INCLUDE THE RETURNED DEVICE (IF AVAILABLE), PHOTOGRAPHS, VIDEOS, EVENT DESCRIPTION, AND ANY ADDITIONAL FIELD INPUT ARTHREX HAS DETERMINED THE MOST LIKELY CAUSE. THE MOST LIKELY CAUSE IS ATTRIBUTED TO USER-RELATED FACTORS, SUCH AS IF THE 3.7 MM TUNNELS THROUGH THE CLAVICLE AND CORACOID ARE NOT PROPERLY ALIGNED, THE IMPLANT MAY NOT SEAT CORRECTLY, PREVENTING EFFECTIVE TENSIONING AND REDUCTION.