inforMED
MalfunctionBRZ

CLEARLINK

Received Feb 3, 2026 · Event occurred Jan 6, 2026

Report 1416980-2026-00464 · MDR key 24251375

Device

Generic name

Set, Blood Transfusion

Model number

2C8750

Catalog number

2C8750

Lot number

DR25C06048

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: B3/D10, D4 EXPIRATION DATE (UPDATE TO N/A), D9, H3, H4, H6 (UPDATE CODES), H11. H11: THE DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED WHICH IDENTIFIED THE SPIKE DETACHED FROM THE TUBE. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION IS RELATED TO ASSEMBLY PROCESS DURING MANUFACTURING. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative

(B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT A CLEARLINK SYSTEM Y-TYPE BLOOD/SOLUTION SET CAME APART DURING PATIENT TRANSFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.