inforMED
InjuryFCD

SUPER XL DELIVERY SYSTEM

Received Sep 27, 1999 · Event occurred Oct 12, 1998

Report 2647856-1999-00040 · MDR key 242442

Device

Generic name

Empty Enema Kit For Air Contrast Studies

Manufacturer

E-z-em, Inc.

Model number

NA

Catalog number

8925

Lot number

12498925

Product problems

  • Loose
  • Increase in Pressure
  • Other (for use when an appropriate device code cannot be identified)
  • Defective Component

Patient

*

  • Internal Organ Perforation

Narrative

Description of Event or Problem

DURING A WATER SOLUBLE BARIUM ENEMA, THERE WAS A PERFORATION OF THE PT'S COLON. DURING A RETROSPECTIVE REVIEW OF THIS INCIDENT, FINALIZED 5/24/99, IT WAS DETERMINED THAT THE ROOT CAUSE OF THE PERFORATION WAS A CLAMP ON THE BARIUM ENEMA BAG THAT DID NOT HOLD IN THE CLOSED POSITION. THE PT'S CONDITION HAS BEEN STABLE SINCE THE INCIDENT.