inforMED
MalfunctionHTJ

DEPTH DEVICE, 2.7 / 3.5 / 4.0MM, LOW PRO

Received Feb 2, 2026 · Event occurred Jan 12, 2026

Report 1220246-2026-00359 · MDR key 24238668

Device

Generic name

Depth Gauge

Manufacturer

Arthrex, Inc.

Catalog number

AR-8943-15

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON 12-JAN-2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-8943-15 DEPTH DEVICE FROM AN RAR-9943S LOANER ANKLE FRACTURE SET SN: 14188044 HAD A BROKEN THREADED RING THAT KEEPS THE STICK IN PLACE. IT WAS REPORTED THAT THEY WERE UNSURE IF THE RING HAPPENED TO BE THE WRONG SIZE OR WAS ACTUALLY BROKEN. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: G3, H3, H6. THE COMPLAINT DEVICE WAS NOT RECEIVED FOR EVALUATION. BASED ON THE INFORMATION PROVIDED, WHICH MAY INCLUDE THE RETURNED DEVICE (IF AVAILABLE), PICTURES, VIDEOS, EVENT DESCRIPTIONS, AND ANY ADDITIONAL FIELD DATA, ARTHREX CONCLUDED THE MOST LIKELY CAUSE OF THE REPORTED FAILURE. THE MOST LIKELY CAUSE OF THE REPORTED FAILURE IS A USER ERROR, INCLUDING DAMAGE TO THE DEVICE DUE TO MISHANDLING. THE COMPLAINT ALLEGATION WAS CONFIRMED BASED ON THE CUSTOMER'S ATTACHED VIDEO, WHICH SHOWS THE DEVICE'S NEEDLE DETACHED FROM THE SHAFT. DAMAGE TO THE DEVICE WILL PREVENT THE DEVICE FROM WORKING AS INTENDED.