MalfunctionMAF

WALLSTENT ENDOPROSTHESIS

Received Jan 29, 2026 · Event occurred Nov 27, 2025

Report 2124215-2026-05244 · MDR key 24210405

Device

Generic name

Catheter, Biliary, Diagnostic

Model number

26280

Catalog number

26280

Lot number

0033045282

Product problems

  • Activation, Positioning or Separation Problem
  • Difficult to Advance
  • Activation Failure

Patient

55 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE STENT PARTIALLY DEPLOYED. THE TARGET LESION WAS LOCATED IN THE ILIAC VEIN. A 16X60/9FR UNI PLUS 75CM WALLSTENT UNI STENT SELF-EXPANDING WAS SELECTED FOR USE. DURING THE PROCEDURE, RESISTANCE WAS ENCOUNTERED WHILE ADVANCING THE STENT, AND ADVANCEMENT WAS NOT SMOOTH. AFTER REACHING THE LESION LOCATION, THE SHAFT WAS PULLED TO DEPLOY THE STENT; HOWEVER, THE STENT DID NOT RESPOND. RESISTANCE OCCURRED WHEN APPROXIMATELY ONE THIRD OF THE STENT HAD BEEN DEPLOYED. THE STENT WAS WITHDRAWN ALONG THE GUIDEWIRE, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

Additional Manufacturer Narrative

E1 - INITIAL REPORTER PHONE: (B)(6).