Received Jan 28, 2026 · Event occurred Dec 16, 2025
Report 3005899764-2026-00007 · MDR key 24203864
Device
Generic name
Evolution Transfer Carriage
Manufacturer
Steris Mexico, S. De R.l. De C.v.Model number
AX083020003Catalog number
AX083020003
Product problems
- Mechanical Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT THE FRONT WHEELS TO THEIR EVOLUTION TRANSFER CARRIAGE DID NOT LATCH PROPERLY TO THE STERILIZER'S DOCKING STATION SUBSEQUENTLY CAUSING THE CARRIAGE TO FALL. NO REPORT OF INJURY.
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE TRANSFER CARRIAGE. THE TECHNICIAN FOUND THAT THE LATCHING MECHANISM FOR THE FRONT WHEELS WAS OUT OF ALIGNMENT SUBSEQUENTLY CAUSING THE TRANSFER CARRIAGE TO NOT LATCH PROPERLY DURING UNLOADING RESULTING IN THE REPORTED EVENT. THE TECHNICIAN ADJUSTED THE LATCHING MECHANISM, TESTED THE FUNCTION AND OPERATION OF THE TRANSFER CARRIAGE AND CONFIRMED IT TO BE OPERATING TO SPECIFICATIONS. THE TRANSFER CARRIAGE WAS RETURNED TO SERVICE, AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED.