Device

Generic name

Intravascular Catheter-snare

Catalog number

ONE1500/B

Lot number

K3221086

Product problems

  • Material Separation
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS RETURNED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED, AND THE MARKER BAND OF THE CATHETER WAS MISSING FROM THE TIP. NO ROOT CAUSE WAS ABLE TO BE CONFIRMED. THE DEVICE HISTORY RECORD WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND.

Additional Manufacturer Narrative

THE SUSPECT DEVICE IS EXPECTED TO RETURN FOR EVALUATION. A FOLLOW UP WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem

THE ACCOUNT ALLEGES THAT WHILE USING A ONE SNARE DEVICE, THE DEVICE'S RADIOPAQUE MARKER DETACHED WHILE THE SNARE WAS IN THE PATIENT. THE DETACHED MARKER WAS UNABLE TO BE RETRIEVED AND ULTIMATELY BECAME EMBEDDED IN THE CORONARY SINUS OF THE PATIENT'S HEART. DUE TO THIS COMPLICATION, THE PATIENT WAS TRANSFERRED TO ANOTHER FACILITY FOR FURTHER EVALUATION.