DERMASENSOR
Received Jan 27, 2026
Report 3017254-2025-00004 · MDR key 24195965
Device
Generic name
Software-aided Adjunctive Diagnostic Device For Use By Physicians On Lesions Sus
Manufacturer
Dermasensor IncModel number
10101Catalog number
10101
Product problems
- Nonstandard Device
- Product Quality Problem
- Manufacturing, Packaging or Shipping Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.
Remedial action
- Recall