Additional Manufacturer Narrative
THE MANUFACTURER PREVIOUSLY REPORTED INFORMATION RELATED TO AN ALICE NIGHTONE SLEEP DIAGNOSTICS DEVICE. THE CUSTOMER COMPLAINT WAS UNIT IS NOT RECORDING PROPERLY. THE DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER. THE SERVICE TECHNICIAN VISUALLY INSPECTED THE DEVICE. THE TECHNICIAN REPORTS DAMAGES TO THE SPO2 CABLE (CABLE INSULATION DAMAGE, WIRES EXPOSED, SPLITTING WIRES EXPOSED, CUT CABLE). CORRECTION: THIS EVENT IS NO LONGER CONSIDERED REPORTABLE BASED ON THE RISK ASSESSMENT.
Description of Event or Problem
THE MANUFACTURER RECEIVED INFORMATION RELATED TO AN ALICE NIGHTONE SLEEP DIAGNOSTICS DEVICE. THE CUSTOMER COMPLAINT WAS UNIT IS NOT RECORDING PROPERLY. THE DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER. THE SERVICE TECHNICIAN VISUALLY INSPECTED THE DEVICE. THE TECHNICIAN REPORTS DAMAGES TO THE SPO2 CABLE (CABLE INSULATION DAMAGE, WIRES EXPOSED, SPLITTING WIRES EXPOSED, CUT CABLE). A FINAL REPORT IS BEING SUBMITTED. IF NEW INFORMATION BECOMES AVAILABLE, A FOLLOW UP/SUPPLEMENTAL REPORT MAY BE COMPLETED.