inforMED
MalfunctionJOS

BIPOLAR CUTTING LOOP ELECTRODE, ROUND

Received Dec 12, 2011 · Event occurred Oct 17, 2011

Report 1418479-2011-00022 · MDR key 2418090

Device

Generic name

Bipolar Cutting Electrode

Model number

4622.1313

Catalog number

4622.1313

Product problems

  • Break
  • Detachment Of Device Component

Patient

29 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE FOLLOWING IS REPORTED TO RWMIC: EVENT DESCRIPTION: DURING THE PROCEDURE, THE BIPOLAR WOLF RESECTOSCOPE 26 FRENCH LOOP ELECTRODE BROKE OFF INSIDE THE UTERUS. MD, USING ANOTHER RESECTOSCOPE WAS ABLE TO RETRIEVE THE SMALL LOOP THAT HAD BROKEN OFF IN THE UTERINE CAVITY, WITHOUT INJURY TO THE PT. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? HYSTEROSCOPIC MYOMECTOMY. DEVICE USAGE PROBLEM: DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING).

Additional Manufacturer Narrative

DEVICE EVALUATED BY MFR: RA NUMBER WAS PROVIDED TO FACILITY, DEVICE HAS NOT BEEN RECEIVED AS OF (B)(4) 2011. SUSPECT MEDICAL DEVICE: PART NUMBER WAS ORIGINALLY REPORTED AS 4622133, THIS WAS IN ERROR. RWMIC WILL VERIFY PART NUMBER AND LOT NUMBER WHEN ACTUAL PRODUCT IS RECEIVED.