inforMED
MalfunctionFMT

GIRAFFE OMNIBED CARESTATION

Received Jan 26, 2026 · Event occurred Dec 29, 2025

Report 2112667-2026-00516 · MDR key 24176674

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN O2 DELIVERY FAILURE. THERE WAS NO PATIENT INJURY.

Additional Manufacturer Narrative

GE HEALTHCARES INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: PATIENT INFORMATION COULD NOT BE OBTAINED DUE TO COUNTRY PRIVACY LAWS.  LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE CABLES FROM THE OXYGEN SERVO BOARD WERE RECONNECTED TO RESOLVE THE ISSUE.