Received Jan 25, 2026 · Event occurred Dec 12, 2025

Report 1528319-2026-00003 · MDR key 24175832

Device

Generic name

Raptor Grasping Device

Manufacturer

Us Endoscopy

Model number

00711177

Catalog number

00711177

Product problems

  • Mechanical Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THE JAWS TO THEIR RAPTOR GRASPING DEVICE "BROKE AND WOULD NOT CLOSE" DURING A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

Additional Manufacturer Narrative

THE RAPTOR GRASPING DEVICE SUBJECT OF THE EVENT WAS DISCARDED BY USER FACILITY PERSONNEL PRIOR TO BEING RETURNED FOR EVALUATION. WITHOUT THE RETURN OF THE DEVICE, A ROOT CAUSE COULD NOT BE DETERMINED. THE RAPTOR GRASPING DEVICE INSTRUCTIONS FOR USE STATES, "ACTUATE THE DEVICE BY MOVING THE SLIDER ON THE HANDLE BACK AND FORTH TO CONFIRM THAT THE GRASPING JAWS OPEN AND CLOSE SMOOTHLY. IF THE UNIT DOES NOT FUNCTION PROPERLY, OR THERE IS EVIDENCE OF DAMAGE (E.G. BENDS, KINKS, MISSHAPEN JAWS, MISALIGNED JAWS, EXPOSED WIRES) DO NOT USE THIS PRODUCT AND CONTACT YOUR LOCAL PRODUCT SPECIALIST. DO NOT USE EXCESSIVE FORCE ON THE HANDLE AND DO NOT COIL THE CATHETER OUTSIDE OF THE ENDOSCOPE. EXCESSIVE FORCE OR COILING MAY DAMAGE THE DEVICE OR DAMAGE THE ENDOSCOPE AND MAY RESULT IN ACCIDENTAL INJURY TO THE PATIENT OR CLINICIAN. THE FOLLOWING CONDITIONS MAY CAUSE THE DEVICE TO FUNCTION IMPROPERLY: ADVANCING THE HANDLE TO THE OPEN POSITION WITH TOO MUCH SPEED OR FORCE. ATTEMPTING TO PASS OR OPEN THE DEVICE IN AN EXTREMELY ARTICULATED ENDOSCOPE. ATTEMPTING TO ACTUATE THE DEVICE IN AN EXTREMELY COILED POSITION. ACTUATING THE DEVICE WHEN THE HANDLE IS AT AN ACUTE ANGLE IN RELATION TO THE SHEATH." A STERIS ACCOUNT MANAGER OFFERED IN-SERVICE TRAINING ON THE PROPER USE AND OPERATION OF THE RAPTOR GRASPING DEVICE; HOWEVER, THE USER FACILITY DECLINED. NO ADDITIONAL ISSUES HAVE BEEN REPORTED. UDI RELATED DATA CLARIFICATION - THE LOT/SERIAL NUMBER IS UNKNOWN, THEREFORE, THE APPLICABLE PRODUCTION IDENTIFIERS WERE NOT INCLUDED IN THE MDR.