inforMED
MalfunctionHSB

PRECICE® BONE TRANSPORT SYSTEM

Received Jan 23, 2026 · Event occurred Dec 23, 2025

Report 3006179046-2026-00101 · MDR key 24167402

Device

Generic name

Rod, Fixation, Intramedullary And Accessories

Model number

BT10-10D380-7

Lot number

3081003 AAA

Product problems

  • Failure to Advance

Patient

40 YR · Female

  • Failure of Implant

Narrative

Additional Manufacturer Narrative

THE DEVICE HAS NOT YET BEEN RECEIVED FOR EVALUATION. THE ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL INFORMATION.

Description of Event or Problem

IT WAS REPORTED THAT AFTER THE INITIAL OR DATE ON (B)(6) 2025 WHERE THE ANTEGRADE IMPLANTATION WAS PERFORMED WITH RETROGRADE TRANSPORT BY USING A CABLE WIRE. THE SLOT WAS MOVED DISTALLY USING FDM. A SURGICAL TOWEL WAS PLACED BETWEEN NAIL AND FDM. THE MANEUVER FUNCTIONED WITHOUT DIFFICULTY. THE IMPLANTATION OF THE NAIL PROCEEDED AS EXPECTED, MARKING THE MAGNET. THEN ON (B)(6) 2025 THERE WAS THE PATIENT ERC4 INTRODUCTION. WE RENEWED CONTACT ON (B)(6) 2025 WHERE THE SURGEON NOTICED THE INCOMPLETE OSTEOTOMY. ON (B)(6) 2025 WE HAD THE REVISION PROCEDURE. THIS INCLUDED THE COMPLETION OF THE OSTEOTOMY. AND FOLLOWING COMPLETION OF THE OSTEOTOMY, THE FAST DISTRACTOR MAX WAS APPLIED IN AN ATTEMPT TO MOVE THE SLOT. THE DEVICE WAS MAINTAINED AT THE MARKED POSITION FOR A MINIMUM OF 5 MINUTES UNDER XRAY CONTROL. NO MOVEMENT OF THE SLOT WAS OBSERVED. SUBSEQUENTLY, THE ERC 4 WAS APPLIED IN AN ADDITIONAL ATTEMPT TO ACTIVE THE SLOT MOVEMENT. AGAIN, NO MOVEMENT WAS DETECTED. THE CABLE WIRE WAS REMOVED. BASED ON INTRAOPERATIVE FINDINGS, NAIL EXCHANGE WAS DEEMED NOT SURGICALLY FEASIBLE. THE BONE TRANSPORT WILL NOW BE PERFORMED USING AN EXTERNAL RING FIXATOR. ADDITIONALLY, REGARDING REVISION SURGERY WE CHECKED THE ERC4 HISTORY, NAIL AND ERC CONNECTED AT EVERY TIME, SINCE THE TRANSPORT WERE STARTED FROM (B)(6) 2025 WITH 3X 0.33MM PER DAY TILL (B)(6) 2025.

Description of Event or Problem

IT WAS REPORTED THAT AFTER THE INITIAL OR DATE ON (B)(6) 2025 WHERE THE ANTEGRADE IMPLANTATION WAS PERFORMED WITH RETROGRADE TRANSPORT BY USING A CABLE WIRE. THE SLOT WAS MOVED DISTALLY USING FDM. A SURGICAL TOWEL WAS PLACED BETWEEN NAIL AND FDM. THE MANEUVER FUNCTIONED WITHOUT DIFFICULTY. THE IMPLANTATION OF THE NAIL PROCEEDED AS EXPECTED, MARKING THE MAGNET. THEN ON (B)(6) 2025 THERE WAS THE PATIENT ERC4 INTRODUCTION. WE RENEWED CONTACT ON (B)(6) 2025 WHERE THE SURGEON NOTICED THE INCOMPLETE OSTEOTOMY. ON (B)(6) 2025 WE HAD THE REVISION PROCEDURE. THIS INCLUDED THE COMPLETION OF THE OSTEOTOMY. AND FOLLOWING COMPLETION OF THE OSTEOTOMY, THE FAST DISTRACTOR MAX WAS APPLIED IN AN ATTEMPT TO MOVE THE SLOT. THE DEVICE WAS MAINTAINED AT THE MARKED POSITION FOR A MINIMUM OF 5 MINUTES UNDER XRAY CONTROL. NO MOVEMENT OF THE SLOT WAS OBSERVED. SUBSEQUENTLY, THE ERC 4 WAS APPLIED IN AN ADDITIONAL ATTEMPT TO ACTIVE THE SLOT MOVEMENT. AGAIN, NO MOVEMENT WAS DETECTED. THE CABLE WIRE WAS REMOVED. BASED ON INTRAOPERATIVE FINDINGS, NAIL EXCHANGE WAS DEEMED NOT SURGICALLY FEASIBLE. THE BONE TRANSPORT WILL NOW BE PERFORMED USING AN EXTERNAL RING FIXATOR. ADDITIONALLY, REGARDING REVISION SURGERY WE CHECKED THE ERC4 HISTORY, NAIL AND ERC CONNECTED AT EVERY TIME, SINCE THE TRANSPORT WERE STARTED FROM (B)(6) 2025 WITH 3X 0.33MM PER DAY TILL (B)(6) 2025.

Additional Manufacturer Narrative

THE INVESTIGATION REVEALED THAT A PRECICE BONE TRANSPORT ANTEGRADE FEMUR TROCHANTER, 10MM X 380MM NAIL WAS REPORTED AS FAILING TO DISTRACT AFTER THE NAIL WAS IMPLANTED. THE DISTRACTION ROD WAS FULLY RETRACTED DISTALLY, AND DURING A FOLLOW UP APPOINTMENT, IT WAS NOTICED THAT THERE WAS AN INCOMPLETE OSTEOTOMY. DURING THE REVISION SURGERY, AND AFTER THE OSTEOTOMY WAS COMPLETE, A FAST DISTRACTOR MAX WAS USED IN AN ATTEMPT TO MOVE THE DISTRACTION ROD PROXIMALLY, BUT THE DISTRACTION ROD WOULD NOT MOVE. THE NAIL REMAINS IN-SITU, AND THE TRANSPORT WILL BE FINISHED WITH THE USE OF A RING FIXATOR. THE DEVICE WAS NOT MADE AVAILABLE FOR EVALUATION. REVIEWING THE PROVIDED X-RAYS OF THE INCOMPLETE OSTEOTOMY, THERE DOES NOT APPEAR TO BE ANY TENSION IN THE CABLE, WHICH AFTER APPROXIMATELY 20 DAYS OF DISTRACTION WITH NO MOVEMENT OF THE INTERCALARY SEGMENT, THERE SHOULD BE TENSION SHOWN IN THE CABLE. IF THE DISTRACTION ROD WAS FUNCTIONING AS INTENDED, THE INTERCALARY SEGMENT WOULD EITHER BE MOVING, AND THE SAME AMOUNT OF TENSION WOULD BE PRESENT IN THE CABLE AS THE SEGMENT MOVED PROXIMALLY, OR THE SEGMENT WOULD NOT BE MOVING DUE TO THE INCOMPLETE OSTEOTOMY, AND THE CABLE WOULD PRESENT A LARGE AMOUNT OF TENSION. BECAUSE NEITHER IS OCCURRING, WHAT MOST LIKELY OCCURRED IS THAT WHEN THE DISTRACTION ROD WAS RETRACTED DISTALLY, THE NAIL BECAME JAMMED WHICH CAN OCCUR WITH IMPROPER USE OF THE FAST DISTRACTOR MAX OR ERC. THEN WITH THE NAIL NOT MOVING, THE OSTEOTOMY WOULD BEGIN TO HEAL, CREATING THE LOOK OF AN INCOMPLETE OSTEOTOMY. EVEN WITH THE OSTEOTOMY REVISION, THE NAIL WOULD STILL NOT BE ABLE TO MOVE IF THE NAIL WAS JAMMED, WHICH MATCHES THE REPORTED EVENT. IN CONCLUSION, BECAUSE THE NAIL WAS NOT RETURNED, THE COMPLAINT OF A FAILURE TO DISTRACT CANNOT BE CONFIRMED. ADDITIONALLY, THE CAUSE TO THE REPORTED FAILURE IS UNKNOWN. HOWEVER, BASED ON THE PROVIDED X-RAYS, IT'S POSSIBLE THE NAIL JAMMED PRIOR TO IMPLANTATION AFTER BEING RETRACTED WITH A FAST DISTRACTOR MAX. A REVIEW OF THE DHR DOCUMENTS INDICATES THE DEVICE WAS MANUFACTURED BY THE SPECIFIED REQUIREMENT AT THE TIME AND MET ALL OF THE REQUIRED INSPECTIONS BEFORE SHIPMENT. THE ASSOCIATED RISK DOCUMENT WAS REVIEWED, AND THE FAILURE MODE, EFFECTS, AND HARMS POTENTIALLY RELATED TO THE REPORTED EVENT HAVE BEEN IDENTIFIED AND MITIGATED. FURTHERMORE, THE COMPLAINT RATE IS WITHIN THE ESTIMATED RATE IN THE RISK DOCUMENTATION ASSOCIATED WITH THIS PRODUCT. BASED ON THIS RISK ASSESSMENT, NO NEW OR ADDITIONAL RISK EVALUATION IS NECESSARY.