inforMED
MalfunctionNGT

SYR 10ML PUMP COMPATIBLE SALINE 10ML FIL

Received Jan 22, 2026 · Event occurred Jan 14, 2026

Report 1911916-2026-00024 · MDR key 24145016

Device

Generic name

Saline, Vascular Access Flush

Catalog number

306547

Lot number

5262073

Product problems

  • Infusion or Flow Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT IS REPORTED PUMP COMPATIBILITY ISSUES. VERBATIM: EVENT DESCRIPTION: WHEN USING THESE SALINE FLUSHES, THE PUMPS WILL CONTINUOUSLY BEEP OCCLUDED. OUTCOME: INFUSION INTERRUPTIONS, NEED TO REPLACE SALINE.

Additional Manufacturer Narrative

(B)(4). FOLLOW UP FOR DEVICE EVALUATION: IT WAS REPORTED THAT, WHEN USING THESE SALINE FLUSHES, INFUSION PUMPS CONTINUOUSLY ALARM WITH AN OCCLUSION ALERT. TO SUPPORT THE INVESTIGATION, FOUR HUNDRED TWENTY SAMPLES IN SEALED BLISTER PACKAGING WERE RECEIVED FOR EVALUATION BY THE QUALITY TEAM. FIFTY SAMPLES WERE RANDOMLY SELECTED FOR TESTING. VISUAL INSPECTION DID NOT IDENTIFY ANY DEFECTS OR IMPERFECTIONS. EACH SELECTED SAMPLE WAS THEN EVALUATED FOR SUSTAINING FORCE, DEFINED AS THE FORCE APPLIED TO THE PLUNGER ROD DURING DOWNWARD TRAVEL WHILE EXPELLING THE SALINE SOLUTION, AND ALL RESULTS MET SPECIFICATION REQUIREMENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR MATERIAL NUMBER 306547, LOT 5262073. THE REVIEW DID NOT IDENTIFY ANY QUALITY ISSUES DURING PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. NO RELATED QUALITY NOTIFICATIONS WERE IDENTIFIED. ALL MANUFACTURING PROCESSES AND FINAL INSPECTIONS WERE PERFORMED IN ACCORDANCE WITH ESTABLISHED SPECIFICATIONS AND ACCEPTANCE CRITERIA. BASED ON THE INVESTIGATION AND THE RETURNED SAMPLE EVALUATION, THE CUSTOMER REPORTED SYMPTOM COULD NOT BE CONFIRMED.