inforMED
MalfunctionGWG

CLARUS DIGITAL CHANNELSCOPE ENDOSCOPE, 13.0CM

Received Jan 21, 2026 · Event occurred Dec 3, 2025

Report 24129812 · MDR key 24129812

Device

Generic name

Endoscope, Neurological

Model number

2233-513

Catalog number

2233-513

Lot number

520204

Patient

NA · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

DURING AN ENDOSCOPIC THIRD VENTRICULOSTOMY, DR. WAS USING THE CHANNEL SCOPE IN THE PATIENTS BRAIN FOR VISUALIZATION, WHEN THE SCREEN SUDDENLY WENT BLACK. THE SCOPE WAS IMMEDIATELY REMOVED FROM THE BRAIN, AND WHEN THE NURSE EXAMINED THE CLARUS BOX THE SCOPE WAS PLUGGED INTO, THEY DISCOVERED THERE WAS WATER INSIDE THE CONNECTION SITE. THE TEAM REALIZED WATER FROM THE SCOPE IRRIGATION WAS FLOWING DOWN THE CORD, INTO THE PLUG, CAUSING THE BOX TO SHORT OUT, AND THE SCREEN TO GO BLACK. THE CHANNEL SCOPE WAS REMOVED FROM THE FIELD, AND A DIFFERENT SCOPE SYSTEM WAS USED TO PROCEED WITH THE CASE.